Welcome to PT Pelita Sentosa Jaya. Your trust is our priority

OPTI CCA-TS2 Blood Gas and Electrolyte Analyzer

Provides fast accurate results for blood gases, acid-base balance, electrolytes, ionized calcium, glucose, BUN and lactate.

Runs samples of whole blood, plasma and serum on single-use disposable cassettes.

Uses innovative optical technology.

Requires minimal maintenance.

Description

OPTI CCA-TS2 Blood Gas and Electrolyte Analyzer

  • Provides fast accurate results for blood gases, acid-base balance, electrolytes, ionized calcium, glucose, BUN and lactate.
  • Runs samples of whole blood, plasma and serum on single-use disposable cassettes.
  • Uses innovative optical technology
  • Requires minimal maintenance.

 

Benefits
Innovative, economical and comprehensive

Innovative optical technology

The OPTI CCA-TS2 Blood Gas and Electrolyte Analyzers use OPTI Medical’s optical fluorescence technology. This unique, patented technology does not use electrodes or contact points, which eliminates the need for costly electrode maintenance.

 

Significant cost savings

The OPTI CCA-TS2 analyzers provide lab-quality results with zero standby cost. The single-use consumable is only used during a measurement and safely locks waste inside.

Wide operating temperature

The OPTI CCA-TS2 analyzers have a measuring chamber heated to 37°C (98.6°F). This allows for a wide operating temperature range to meet the needs of colder environments, such as the operating room.

Comprehensive test menu

To help meet exact testing needs, the OPTI CCA-TS2 analyzers have a variety of cassette configurations available. Each cassette comes individually wrapped, in cartons of 25, reducing the potential for consumable waste.

Test menu

Ready-to-use single-use cassettes

The OPTI CCA/CCA-TS Blood Gas and Electrolyte Analyzers use ready-to-use single-use cassettes that let you test for a range of blood gas parameters: Available in Indonesia E-Ca, E-Cl, B-Lac, and E-Lyte CCA

OPTI CCA-TS2 analyzer specifications

Calibration/Measurement

Sample Volume 125 µL (60µL for B60 cassette)
Calibration Time <90 seconds
Measurement Time <120 seconds
Typical Cycle Time <210 seconds
Operating Temperature 10–30°C (50–86°F)
Sample Type Whole blood, serum or plasma
Measurement Principle Optical fluorescence, reflectance
Sample Application Syringe, capillary or ComfortSampler
Data Storage
  • 500 patient records
  • 70 QC samples per level
  • 105 SRC samples per level
  • 300 user IDs and passwords

 

Security

Password Multiple levels of security with user-definable permissions
QC Lockout User-definable QC lockout functions

 

Dimensions/Weight

Height x Width x Depth 124 x 362 x 230 mm (4.7 x 14.2 x 9.1 in)
Weight 4.3 kg (9.6 lb) with rechargeable battery pack

 

Miscellaneous

Voltage/Frequency 120 or 240 AC, 50–60 Hz
Power Consumption (max.) 110 VA
Interface ASTM, ASCII, POCT protocols via USB or Ethernet
Approvals CE, UL, IVDD, and FDA